跳轉至內容
Merck
全部照片(1)

Key Documents

G005458

HPLC Analysis of mAb Charge Variants on TSKgel® CM-STAT

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: TSKgel CM-STAT, 10 cm x 4.6 mm I.D., 7 μm particles (821966)
mobile phase: [A] 20 mM MES buffer, pH 6.0; [B] 0.5 M NaCl in buffer A, pH 6.0
gradient: 0% B (0 min), 30% B (15 min), 100% B (15 min),

0% B (17 min), 10% B (17 min), 10% B (21 min)

flow rate: 1 mL/min
column temp.: ambient
detector: UV, 280 nm
injection: 20 μL
sample: monoclonal antibodies (mAb A through E)

適合性

application for HPLC

應用

life science and biopharma

分析報告

A TSKgel CM-STAT weak cation exchange (WCX) column was applied to separate charge variants of several monoclonal antibodies. The typical analysis time on conventional 25 cm long WCX columns of about eighty minutes could be significantly reduced when separation was performed on a 10 cm TSKgel CM-STAT column, filled with 7 μm particles. The analysis profiles for five antibodies show that high resolution analysis can be obtained in about 20 minutes analysis time.

引用

From "Ion Exchange Chromatography for the Characterization of Biotherapeutics" in The Supelco Reporter, Vol. 29.3, page 20.

法律資訊

TSKgel is a registered trademark of Tosoh Corporation

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

抱歉,我們目前沒有在線上提供本產品的COA

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務