跳轉至內容
Merck
全部照片(1)

重要文件

G005166

HPLC Analysis of Positional Isomers (Xylenes) on Astec® CYCLOBOND I 2000

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: CYCLOBOND I 2000, 25 cm x 4.6 mm I.D., 5 μm particles (20024AST)
mobile phase: [A] acetonitrile; [B] water; (15:85, A:B)
flow rate: 0.8 mL/min
column temp.: 55 °C
detector: UV, 230 nm
injection: 3 μL
sample: xylene isomers, each compound, 0.1 mg/mL in acetonitrile:water (50:50) (elution order: m, o, p)

適合性

application for HPLC

應用

chemicals and industrial polymers

法律資訊

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CYCLOBOND is a trademark of Sigma-Aldrich Co. LLC

分析物

描述

    m-Xylene, analytical standard
    o-Xylene, analytical standard
    p-Xylene, analytical standard

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 文件 section.

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務