跳轉至內容
Merck
全部照片(1)

重要文件

G004220

GC Analysis of PCBs in Transformer Oil on a 5% Phenyl Column after SPE using Supelclean Sulfoxide

suitable for GC, application for SPE

登入查看組織和合約定價


About This Item

技術

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

測試參數

SPE tube/cartridge: Supelclean Sulfoxide SPE Tube, Glass 6 g/20 mL (55252-U)
sample preparation: SPE (Solid Phase Extraction)
column: 5% phenyl polysiloxane (Supelco equivalent, SLB-5ms, 28471-U)
sample addition: transformer oil spiked with PCBs
detector: QMS (For full method details, please see reference (1))
extraction process: Sample Pre-treatment: paraffin oil was spiked with a Kanechlor PCB mix at the total levels of 3.7 ppm (mg/kg) and diluted with hexane (1:1 v/v). (Commercial insulation oil (Japan Industrial Standard JIS C2320-1999, insulating oil, Class 1-2/4).)
condition: 20 mL acetone; 40 mL of hexane
sample addition: 0.4 mL diluted oil sample
elution: 12 mL hexane (elutes oil interferences); 25 mL hexane (elutes PCBs)
eluate post-treatment: concentrate under nitrogen

適合性

application for SPE

應用

environmental

應用

Sample Pre-treatment:
Commercial insulation oil (Japan Industrial Standard JIS C2320-1999, insulating oil, Class 1-2/4, paraffin oil) was spiked with a Kanechlor PCB mix at the total levels of 3.7 ppm (mg/kg) and diluted with hexane (1:1 v/v).

法律資訊

Supelclean is a trademark of Sigma-Aldrich Co. LLC

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

抱歉,我們目前沒有在線上提供本產品的COA

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

Masahiko Numata et al.
Analytical chemistry, 75(6), 1450-1457 (2003-03-28)
A novel sample extraction technique for polychlorinated biphenyls (PCBs) and organochlorine pesticides (OCPs) analysis using microwave-heating device is developed. In this study, microwave-assisted extraction (MAE) and steam distillation techniques were combined. Desorption of the anatytes from solid matrixes was accelerated

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.