推薦產品
技術
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
測試參數
sample/matrix: 3 g crushed potato chips in 15 mL vial
SPME fiber: DVB/Carboxen/PDMS, 50/30 μm, StableFlex (57328-U)
extraction: headspace, 65 °C, 20 min, with stirring
desorption process: 3 min, 250 °C
column: SPB-1 SULFUR, 30 m × 0.32 mm I.D., 4.0 μm (24158)
inj. temp.: 250 °C
oven: 45 °C (1.5 min) to 250 °C at 12 °C/min, hold 10 min
detector: quadrupole mass spectrometer, m/z = 35-290 at 0.6 sec/scan
carrier gas: helium, 40 cm/sec
injection: Splitless/split, closed 2 min
liner: 0.75 mm liner
sample: 3g crushed potato chips in 15 mL vial
適合性
application for SPME
應用
food and beverages
分析報告
Potato chips can contain up to 50% oil and are subject to rancidity. Aldehydes, particularly hexanal and heptanal, are formed from fatty acids that are cleaved at the double bonds when exposed to ultraviolet light and air. A package of 3-year-old potato chips was analyzed using SPME and compared to a fresh package. The presence of aldehydes indicated that the older chips were rancid. However, due to the age of the chips, further oxidation of the aldehydes occurred that resulted in the formation of hexanoic and heptanoic acids. A Carboxen/PDMS/DVB fiber was inserted into the headspace of the potato chip samples at 65 °C for 20 minutes with agitation. The fiber was then analyzed using GC/MS on a SPB-1 SULFUR column. This non-polar column has a thick film and is ideal for smaller molecules.
法律資訊
SPB is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
Nonanal
Octane
Butyl hexanoate
Octanoic acid
Heptanoic acid
Hexanal
Methyl butyrate
Heptanal, ≥95%, FCC, FG
Octanal
Decanal
Acetic acid
Hexanoic acid
查看全部 (12)
固相微萃取纖維
標準
產品號碼
描述
訂價
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務