跳轉至內容
Merck
全部照片(1)

Key Documents

795-0567

GC Analysis of Nitrogen-Containing Herbicides in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber

suitable for GC, application for SPME

登入查看組織和合約定價


About This Item

技術

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

測試參數

sample/matrix: 4 mL (water + 100ng/mL each analyte 1g/mL NaCl), pH 2, in 4.6 mL vial
SPME fiber: polyacrylate, 85 μm (57304)
extraction: immersion, ambient temp., 50 min (constant stirring)
desorption process: 5 min, 230 °C
column: 5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 40 °C (5 min) to 100 °C at 30 °C/min, to 275 °C at 5 °C/min
inj. temp.: splitless, 230 °C
detector: MS (Ion Trap Manifold: 250°C; scan range m/z=45-400 at 0.6 sec/scan)
carrier gas: helium, 40 cm/sec, set at 40 °C

適合性

application for SPME

應用

environmental
food and beverages

分析物

描述

    Atrazine
    Prowl®, analytical standard
    Hexazinone
    Oxadiazon
    Simazine
    Terbacil
    Trifluralin
    Bromacil
    Propazine
    Propachlor
    Eptam®, PESTANAL®, analytical standard
    查看全部 (11)

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

抱歉,我們目前沒有在線上提供本產品的COA

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務