03-0196-00
SMC® Human Total α-synuclein High Sensitivity Immunoassay Kit
Uses 100 µL per well of 1:200 diluted serum or plasma samples, or 1:10 diluted cerebrospinal fluid (CSF) samples.
同義詞:
Human α-synuclein, α-synuclein, Human total-α-synuclein, a-synuclein, total-a-synuclein
登入查看組織和合約定價
全部照片(1)
About This Item
物種活性
canine, rat, feline, mouse
製造商/商標名
SMC®
assay range
sensitivity: 1.95 pg/mL
(LLOQ of Total α-synuclein)
inter-assay cv: 14%
intra-assay cv: 9%
(five normal plasma samples run in triplicate by 3 different operators on 3 different days)
技術
single molecule counting: suitable
一般說明
The SMC® Human Total-α-Synuclein High Sensitivity Immunoassay Kit contains all reagents required to perform the assay. This assay takes advantage of a standard immunoassay workflow configured in a standard plate format. Assay protocols are similar to existing sandwich ELISA methods with two key differences:
1) Elution buffer disrupts the sandwich, separating the labeled detection antibody for quantification.
2) The Erenna and SMCxPRO® immunoassay systems detect analytes using Single Molecule Counting (SMC®) technology.
1) Elution buffer disrupts the sandwich, separating the labeled detection antibody for quantification.
2) The Erenna and SMCxPRO® immunoassay systems detect analytes using Single Molecule Counting (SMC®) technology.
特異性
Species Reactivity Note
Specificity to the following species samples were tested with the following results:
Monkey – 100%
Cat – 100%
Rat – 100%
Mouse – 100%
Beagle – 100%
Cross Reactivity
Cross-reactivity to the following analytes were tested with the following results:
Tau – 0%UCHL1 – 0%
β-synuclein – 0%
γ-synuclein – 0%
Accuracy
CSF Recovery: 5 CSF samples spiked with average recovery of 96% and a range of 90-102%
Serum Recovery: 4 serum samples spiked with average recovery of 96% and a range of 93-98%
Plasma Recovery: 4 plasma samples spiked with average recovery of 96% and a range of 93-97%
Specificity to the following species samples were tested with the following results:
Monkey – 100%
Cat – 100%
Rat – 100%
Mouse – 100%
Beagle – 100%
Cross Reactivity
Cross-reactivity to the following analytes were tested with the following results:
Tau – 0%UCHL1 – 0%
β-synuclein – 0%
γ-synuclein – 0%
Accuracy
CSF Recovery: 5 CSF samples spiked with average recovery of 96% and a range of 90-102%
Serum Recovery: 4 serum samples spiked with average recovery of 96% and a range of 93-98%
Plasma Recovery: 4 plasma samples spiked with average recovery of 96% and a range of 93-97%
應用
Uses 100 µL per well of 1:200 diluted serum or plasma samples, or 1:10 diluted cerebrospinal fluid (CSF) samples.
包裝
96 well assay
成分
- Assay Buffer
- Total α-synuclein Coated Beads
- Standard Diluent
- Total α-synuclein Detection Antibody
- Total α-synuclein Standard
- 10X Wash Buffer
- Buffer D
- Elution Buffer B
其他說明
- This new streamline kit has the following features:
- Improved options for sample prep and incubation
- Lyophilization protein standard for uniform shipping and storage condition
- Increased reagent volumes to accommodate automation
- Able to achieve assay specifications on both SMCxPRO® and Erenna.
法律資訊
SMC is a registered trademark of Singulex
SMCxPRO is a registered trademark of Merck KGaA, Darmstadt, Germany
免責聲明
Unless otherwise stated in our catalog or other company documentation accompanying the product(s), our products are intended for research use only and are not to be used for any other purpose, which includes but is not limited to, unauthorized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeutic uses or any type of consumption or application to humans or animals.
訊號詞
Danger
危險分類
Acute Tox. 3 Dermal - Acute Tox. 4 Inhalation - Acute Tox. 4 Oral - Aquatic Chronic 3 - Eye Irrit. 2 - Repr. 1B
儲存類別代碼
6.1D - Non-combustible acute toxic Cat.3 / toxic hazardous materials or hazardous materials causing chronic effects
閃點(°F)
Not applicable
閃點(°C)
Not applicable
分析證明 (COA)
輸入產品批次/批號來搜索 分析證明 (COA)。在產品’s標籤上找到批次和批號,寫有 ‘Lot’或‘Batch’.。
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務