技術
HPLC: suitable
測試參數
sample/matrix: Plasma proteins were removed with addition of 900 μL of acetonitrile, containing deuterated internal standards, to 250 μL of human EDTA plasma. The mixture was vortexed, (centrifuged and the supermatant evaporated before being reconstituted in a 50:50 solution of methanol and water. 10 μL (corresponding to 8.57 μL of plasma) was injected into the HPLC.)
column: Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)
mobile phase: [A]: Water (5 mM ammonium acetate and 0.012% formic acid); [B]: Methanol (5 mM ammonium acetate and 0.012% formic acid)
gradient: 70 to 95% B in 10 minutes held 4 minutes
flow rate: 0.6 mL/min
column temp.: 40 °C
detector: ESI(-), MRM Mode
適合性
application for HPLC
應用
clinical
分析報告
法律資訊
分析物
分析柱
標準品
流動相成分