Skip to Content
Merck
All Photos(1)

Key Documents

G005231

US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm), Fast GC Analysis

suitable for GC

Sign Into View Organizational & Contract Pricing


About This Item

technique(s)

gas chromatography (GC): suitable

test parameters

column: SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 50 °C (2 min), 20 °C/min to 325 °C (1 min)
inj. temp.: 250 °C
detector: FID, 350 °C
carrier gas: helium, 0.9 mL/min, constant
injection: 1 μL, splitless (splitter open at 0.5 min)
liner: 2 mm I.D., FocusLiner inlet liner
sample: each analyte at 50 μg/mL in methylene chloride

application(s)

environmental

Legal Information

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    Decane
    Dodecane
    Tetradecane
    Hexadecane
    Octadecane
    Eicosane
    Docosane
    Tetracosane
    Hexacosane
    Octacosane
    See All (10)

recommended

Product No.
Description
Pricing

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Customer Support.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service