G001056
HPLC Analysis of Barbiturates in Serum on Discovery® C18 after SPE using Discovery® DSC-18Lt
application for SPE, application for HPLC
About This Item
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 0.5 mL porcine serum spiked with 0.5 μg/mL or 1.0 μg/mL each analyte then diluted with 0.5 mL water
SPE tube/cartridge: Discovery DSC-18Lt, 500 mg/3 mL (52613-U)
condition: 2 mL methanol; 2 mL DI water
sample addition: 1 mL at 0.75 mL/min
washing: 2 mL 5% methanol, then vacuum or air dry for 5-10 min
elution: 1-2 mL methanol
eluate post-treatment: dry eluate with nitrogen purge (40 °C; 15-20 min), reconstitute in 20 μL mobile phase
column: Discovery C18, 15 cm × 4.6 mm, 5 μm preceded by a 2 cm C18 guard column and 0.5 μm frit filter (504955)
mobile phase: [A] methanol; [B] water (40:60, A:B)
flow rate: 1 mL/min
column temp.: 30 °C
detector: UV, 214 nm
injection: 30 μL, diluted porcine serum extract
suitability
application for HPLC
application for SPE
application(s)
forensics and toxicology
Analysis Note
Legal Information
Analyte
SPE tube or plate
analytical column
standard
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service