Skip to Content
Merck
All Photos(1)

Key Documents

G003758

GC Analysis of Pesticides in Spinach on Equity-1 after SPE using Supelclean ENVI-Carb-II/PSA

suitable for GC, application for SPE

Sign Into View Organizational & Contract Pricing


About This Item

technique(s)

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

test parameters

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 10 g of fresh spinach spiked at 0.2 ppm each pesticide
SPE tube/cartridge: Supelclean ENVI-Carb-II/PSA, 500mg/300mg/6mL (55119-U)
condition: 5 mL acetonitrile:toluene (3:1, v/v)
sample addition: 1 mL concentrated sample extract
elution: 20 mL acetonitrile:toluene (3:1, v/v)
eluate post-treatment: Evaporate eluate to dryness with nitrogen at 40 °C. Reconstitute in 1 mL acetone:hexane (1:1, v/v).
column: Equity-1, 30 m x 0.25 mm I.D., 0.25 μm (28046-U)
oven: 50 °C (5 min), 25 °C/min to 125 °C, 10 °C/min to 30 °C (8 min)
inj. temp.: 200 °C, aux: 325 °C (8 min)
detector: MSD, scan range 45-450 amu
carrier gas: helium, 0.9 mL/min., constant flow mode
injection: 1 μL, splitless (splitter open at 1 min)
liner: 4 mm I.D., single taper

suitability

application for SPE

application(s)

agriculture
food and beverages

Application

Sample Pre-treatment:
Homogenize 10 g of spinach in 10 mL of acetonitrile.

Legal Information

Supelclean is a trademark of Sigma-Aldrich Co. LLC

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Customer Support.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service