G005938
GC Analysis of Volatiles in Cumin Powder on SPB®-HAP after Collection/Desorption using Carbotrap® 300 Tube
suitable for GC, application for air monitoring
About This Item
technique(s)
gas chromatography (GC): suitable
test parameters
sample/matrix: 0.005 g Cumin Powder
dry purge: 2 min
desorption process: 350 °C with 10 min hold
transfer line/valve temp.: 180 °C
column: SPB®-HAP, 60 m x 0.32 mm I.D., 4 μm (25020-U)
oven: 40 °C (3 min), 12 °C/min to 200 °C (20 min)
detector: MSD
MSD interface: 210 °C
carrier gas: helium
thermal desorption tube: Carbotrap® 300 Glass-Fritted Thermal Desorption Tube, 1/4 in O.D. x 3.5 in L
suitability
application for air monitoring
application(s)
air monitoring
flavors and fragrances
food and beverages
Analysis Note
Legal Information
Analito
Related product
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico