G005849
LC/MS/MS Analysis of Warfarin Enantiomers in Plasma on Astec® CHIROBIOTIC® V after SPE using HybridSPE®-Phospholipid
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
sample/matrix: rabbit plasma, unfiltered K2-EDTA, spiked with warfarin at 100 ng/mL (3:1, plasma:1% formic acid in acetonitrile)
SPE tube/cartridge: HybridSPE-Phospholipid, 96-well plate (575656-U)
sample addition: to each well add 100 μL plasma, followed by a 300 μL of 1% formic acid in acetonitrile, agitate on orbital shaker for 4 minutes
elution: collect filtrate and analyze directly
column: Astec CHIROBIOTIC® V, 10 cm x 4.6 mm I.D., 5 μm particles (11022AST)
mobile phase: [A] 0.1% formic acid (pH unadjusted); [B] acetonitrile, 75:25 (A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: ESI+, 100-1000 m/z
injection: 2 μL
suitability
application for HPLC
application(s)
clinical
pharmaceutical (small molecule)
Analysis Note
Legal Information
Analito
SPE tube or plate
Standard
analytical column
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico