HPLC Analysis of Levothyroxine and Liothyronine on Ascentis® Express ES-Cyano

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express ES-Cyano, 10 cm x 2.1 mm I.D., 2.7 μm particles (53491-U)
mobile phase: [A] water with 0.05% phosphoric acid; [B] acetonitrile with 0.05% phophoric acid; (60:40, A:B)
flow rate: 1.5 mL/min
pressure: 3270 psi (225 bar)
column temp.: 30 °C
detector: UV, 225 nm
injection: 5 μL
sample: 20 μg/mL in mobile phase

suitability

application for HPLC

application(s)

clinical

Analysis Note

This application demonstrates the suitabilty of Ascentis Express ES-Cyano for the analysis of Levothyroxine according to USP Monograph 3301 with Liothyronine as a reference standard.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descripción

    3,3′,5-Triiodo-L-thyronine sodium salt, powder, BioReagent, suitable for cell culture
    Levothyroxine sodium hydrate, PESTANAL®, analytical standard

analytical column

Referencia del producto
Descripción
Precios

standard

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico