G003765
HPLC Analysis of Hydrocortisone in Topical Cream on Discovery® HS C18 after SPE using Discovery® DSC-Si
application for SPE, application for HPLC
About This Item
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample addition: hydrocortisone topical hand cream in hexane:ethylacetate
SPE tube/cartridge: Discovery DSC-Si, 500 mg/3 mL (52695-U)
condition: 1 mL hexane:ethyl acetate (2:1, v/v)
sample addition: 1 mL
washing: 1 mL hexane:ethyl acetate (2:1, v/v)
elution: 1 mL methanol:DI water (1:1, v/v)
eluate post-treatment: dilute eluate to 2 mL with methanol:DI water (1:1, v/v)
column: Discovery HS C18, 15 cm x 4.6 mm I.D., 5 μm particles (568520-U)
mobile phase: [A] methanol; [B] DI water (1:1, A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: UV, 220 nm
injection: 10 μL
suitability
application for HPLC
application for SPE
application(s)
cleaning products
cosmetics
personal care
pharmaceutical (small molecule)
Application
Dissove 1 g 1% hydrocortisone topical hand cream in 10 mL ethyl acetate. Dilute 2 mL of ethyl acetate-cream sample with 8 mL hexane:ethyl acetate (2:1, v/v).
Legal Information
Analytical column
SPE tube or plate
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico