1.51015
TLC aluminum sheets, silica gel 60 RP-18 F254s MS-grade
pkg of 20 sheets, sheet L × W 5 cm × 7.5 cm, aluminum support
Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización
About This Item
Productos recomendados
material
aluminum support
silica gel 60 matrix
Quality Level
feature
binder Organic Polymer
fluorescent indicator
packaging
pkg of 20 sheets
technique(s)
TLC-MS: suitable
thin layer chromatography (TLC): suitable
layer thickness
100 μm
sheet L × W
5 cm × 7.5 cm
particle size
10-12 μm
pore size
60 Å medium pore diameter
storage temp.
2-30°C
General description
20 Aluminium sheets 5x7.5
Reference:
Gertrud Morlock, Wolfgang Schwack
Coupling of planar chromatography to mass spectrometry
TrAC Trends in Analytical Chemistry, Volume 29, Issue 10, 2010, Pages 1157-1171
Gertrud Morlock, Wolfgang Schwack
Coupling of planar chromatography to mass spectrometry
TrAC Trends in Analytical Chemistry, Volume 29, Issue 10, 2010, Pages 1157-1171
Analysis Note
Specific surface area(according to BET; 5-Pt. measurement): 480 - 540 m²/g
Pore volume(N₂-isotherm): 0.74 - 0.84 ml/g
d 50 (laser diffraction, size distribution):9.7 - 11.7 µm
Layerthickness: 80 - 120 µm
Deviation of layer thickness per plate: ≤ 30 µm
A) Cholesterin test
hRf-values
- cholesterin: 35 - 43
B) Steroid test
- methyltestosterone: 14 - 24
- Reichstein S: 28 - 38- hydrocortisone: 40 - 50
Typical value determined on an activated sheet (120°C, 15 min)
Eluent A) aceton/ water (95/5 v/v)
Eluent B) aceton/ water (60/40 v/v)
C) MS suitability test
- Angiotensin II (signal-to-noiseratio 50ng): ≥ 25
measured with 5000 shots
- Caffeine (signal-to-ratio 5ng): ≥25
Pore volume(N₂-isotherm): 0.74 - 0.84 ml/g
d 50 (laser diffraction, size distribution):9.7 - 11.7 µm
Layerthickness: 80 - 120 µm
Deviation of layer thickness per plate: ≤ 30 µm
A) Cholesterin test
hRf-values
- cholesterin: 35 - 43
B) Steroid test
- methyltestosterone: 14 - 24
- Reichstein S: 28 - 38- hydrocortisone: 40 - 50
Typical value determined on an activated sheet (120°C, 15 min)
Eluent A) aceton/ water (95/5 v/v)
Eluent B) aceton/ water (60/40 v/v)
C) MS suitability test
- Angiotensin II (signal-to-noiseratio 50ng): ≥ 25
measured with 5000 shots
- Caffeine (signal-to-ratio 5ng): ≥25
Certificados de análisis (COA)
Busque Certificados de análisis (COA) introduciendo el número de lote del producto. Los números de lote se encuentran en la etiqueta del producto después de las palabras «Lot» o «Batch»
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico