์ฝ˜ํ…์ธ ๋กœ ๊ฑด๋„ˆ๋›ฐ๊ธฐ
Merck

HPLC Analysis of Pseudoephedrine Enantiomers on Astecยฎ CHIROBIOTICยฎ T

HPLC Analysis of Pseudoephedrine Enantiomers on Astec® CHIROBIOTIC® T application for HPLC

Materials

๋ถ„์„ ์ปฌ๋Ÿผ

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

Astecยฎ CHIROBIOTICยฎ T Chiral HPLC Column

5 μm particle size, L ร— I.D. 25 cm ร— 4.6 mm

ํ‘œ์ค€ ๋ฌผ์งˆ

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

(+)-Pseudoephedrine hydrochloride

≥98%

(โˆ’)-Pseudoephedrine

purum, ≥96.0% (sum of enantiomers, GC)

(1S,2S)-(+)-Pseudoephedrine

98%

(1R,2R)-(โˆ’)-Pseudoephedrine

98%

CONDITIONS

column

CHIROBIOTIC T, 25 cm x 4.6 mm I.D., 5 μm particles (12024AST)

mobile phase

[A] methanol; [B] acetic acid; [C] triethylamine; (100:0.1:0.1, A:B:C)

flow rate

1 mL/min

column temp.

25 ยฐC

detector

UV, 254 nm

injection

10 μL

sample

1 mg/mL in methanol