콘텐츠로 건너뛰기
Merck
모든 사진(1)

주요 문서

G006300

LC/MS Analysis of Aflatoxin M1 in Milk on Ascentis® Express HILIC after SPE using Supel Tox AflaZea, Quantifying Picogram Concentrations

application for SPE, application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable
solid phase extraction (SPE): suitable

검사 파라미터

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: cow milk
SPE tube/cartridge: Supel Tox AlflaZea, 6 mL (55314-U)
extraction process: centrifuge 45 g liquid milk for 60 min at 5000 rpm (RCF 4863 xg), cool at 0 °C for 30 minutes, remove fat layer, transfer middle layer to collection flask,
sample addition: 10 mL, 2 drop/second using vacuum
washing: 2 x 2 mL acetonitrile (collect with sample)
eluate post-treatment: evaporate to dryness at 70 °C using nitrogen
sample preparation: reconstitut in 500 μL 20:80 acetonitrile:water, filter into 750 μL poly vial using PVDF filter, 0.2 μm
column: Ascentis Express C18, 50 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
mobile phase: [A] 0.1% (v/v) formic acid; [B] acetonitrile
gradient: 90% A for 0.5 min; to 10% A in 0.5 min; held at 10% for 3 min; to 90% A in 0.10 min; held at 90% A for 3 min
flow rate: 0.4 mL/min
column temp.: ambient
detector: MS/MS (MRM transitions)
injection: 10 μL

적합성

application for HPLC
application for SPE

응용 분야

food and beverages

분석 메모

Sample pretreatment: centrifuge 45 g liquid milk for 60 min at 5000 rpm (RCF 4863 xg), cool at 0 °C for 30 minutes, remove fat layer, transfer middle layer to collection flask, add 20 mL acetonitrile to 50 mL centrifuge tube, add 20 mL of sample to acetonitrile, add contents of 1 Supel Q non-buffered tube, immedately disperse salt, quicky place on shaker for 5 minutes, centrifuge at 5000 rpm for 20 min, remove top layer (loading solution)
Aflatoxin M1 is the metabolite of Aflatoxin B1 and can be found in milk products when animals have been fed alatoxin-contminated feeds. Aflatoxins are positively associated with hepataocellular carcinomas. Aflattoxin M1 was detected in milk and quantified at concentrations fo 25 pg/mL. Sample preparation included a protein precipitation with solvent partitionng and sample clean-up using Supel Tox AflaZea SPE tubes. Analysis was perfomed via LC-MS using an Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm HPLC column for the separation. Detection and quanitification of aflatoxin M1 at 25 pg/mL satisfies the current regulatory requirements for aflatoxin M1 for EU treaty participants as stated in EC 1881/2006.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC

분석물

설명

    Aflatoxin M1 solution, 0.5 μg/mL in acetonitrile, analytical standard

SPE 튜브 또는 플레이트

표준 물질


가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.