기술
gas chromatography (GC): suitable
검사 파라미터
column: SPB-1 SULFUR, 30 m x 0.32 mm I.D., 4.0 μm (24158)
oven: 35 ºC, 5 ºC/min to 200 ºC
inj. temp.: 275 °C
detector: FID, 300 ºC; FPD, 300 ºC
carrier gas: helium, 20 cm/sec
injection: 1 μL 100:1 split
sample: Gasoline
응용 분야
petroleum
분석 메모
Two issues must be overcome for the analysis of sulfur compounds in fuels. 1) Detection of these compounds in a hydrocarbon sample matrix, and 2) adsorption of these compounds by the column. Sulfur specific detectors, such as a flame photometric detector (FPD), can be used to overcome the first challenge. A low phase ratio column, such as the SPB-1 SULFUR, can be used to overcome the second challenge. This slide shows the analysis of a gasoline sample, in which the column effluent was split to two detectors. The top chromatogram shows the FID response, and the bottom chromatogram shows the FPD response.
법적 정보
SPB is a registered trademark of Merck KGaA, Darmstadt, Germany
가장 최신 버전 중 하나를 선택하세요:
자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..