콘텐츠로 건너뛰기
Merck
모든 사진(1)

Key Documents

G005738

HPLC Analysis of Cosmetic Preservatives on Ascentis® Express RP-Amide

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Ascentis Express RP-Amide, 10 cm x 3 mm I.D., 2.7 μm particles (53918-U)
mobile phase: [A] water; [B] methanol; (60:40, A:B)
flow rate: 0.6 mL/min
pressure: 2900 psi (200 bar)
column temp.: 50 °C
detector: UV, 250 nm
injection: 1 μL
sample: methyl-, benzylparaben (50 mg/L), ethylparaben (30 mg/L), propyl-, butylparaben (100 mg/L), 2-phenoxyethanol (1.5 g/L),1-phenoxy-2-propanol (3 g/L), all in 40% methanol

적합성

application for HPLC

응용 분야

cleaning products
cosmetics
personal care

분석 메모

With the publication Borremans, M. et.al. 2004, Chromatographia 59: 47-53 as a reference, this method is 10% faster, but with superior resolution of critical pairs; in this case, the minimum USP resolution is 2.4.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Ethylene glycol monophenyl ether
    1-Phenoxy-2-propanol
    Methyl 4-hydroxybenzoate
    Ethyl 4-hydroxybenzoate
    Butylparaben
    Benzyl 4-hydroxybenzoate

가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..