기술
LC/MS: suitable
solid phase extraction (SPE): suitable
검사 파라미터
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with clenbuterol enantiomers at 10 ng/mL
SPE well plate: HybridSPE-Precipitation 96-well Plate, 50 mg/well (575656-U)
sample addition: 100 μL spiked rat plasma followed by 300 μL 1% formic acid in acetonitrile. Mix by vortexing the HybridSPE-PPT plate briefly.
elution: apply vacuum
column: CHIROBIOTIC T, 10 cm x 2.1 mm I.D., 5 μm particles (12018AST)
mobile phase: 10 mM ammonium formate in methanol
flow rate: 0.3 mL/min
column temp.: 30 °C
detector: ABI 3200 QT; ESI(+), MRM: 184/104 m/z (phospholipids); 277.2/203.1 m/z (clenbuterol)
injection: 10 μL
적합성
application for LC-MS
application for SPE
응용 분야
clinical
forensics and toxicology
pharmaceutical (small molecule)
애플리케이션
법적 정보
SPE 튜브 또는 플레이트
분석 컬럼
애플리케이션
표준 물질
가장 최신 버전 중 하나를 선택하세요:
Chromatograms
application for HPLCapplication for HPLCapplication for HPLCapplication for HPLC자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..
고객지원팀으로 연락바랍니다.