Passa al contenuto
Merck
Tutte le immagini(1)

Key Documents

G005675

HPLC Analysis of Benzodiazepines on Ascentis® Express Phenyl-Hexyl with Acetonitrile Mobile Phase

application for HPLC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

HPLC: suitable

Parametri analitici

column: Ascentis Express Phenyl-Hexyl, 5 cm x 4.6 mm I.D., 2.7 μm particles (53348-U)
mobile phase: [A] water; [B] acetonitrile; (62:38, A:B)
flow rate: 1.4 mL/min
pressure: 2600 psi (179 bar)
column temp.: 35 °C
detector: UV, 250 nm
injection: 2 μL
sample: 0.1 g/mL of each in 20% methanol

Compatibilità

application for HPLC

applicazioni

clinical
pharmaceutical (small molecule)

Risultati analitici

The Phenyl-Hexyl phase of the Ascentis Express product line displays the best selectivity of this sample; selectivity is markedly affected by choice of organic modifier. The application also demonstrates feasibility of fast LC with Ascentis Express by virtue of low backpressure and high efficiency.

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analita

Descrizione

    Oxazepam
    Alprazolam
    Clonazepam
    Desmethyldiazepam
    Diazepam

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.