G003762
HPLC Analysis of Piroxicam and 2-Aminopyridine in Urine on Discovery® HS F5 after SPE using Discovery® DSC-MCAX
application for SPE, application for HPLC
About This Item
tecniche
HPLC: suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: human urine spiked with piroxicam (4 μg/mL) and 2-aminopyridine (10 μg/mL)
SPE tube/cartridge: Discovery DSC-MCAX, 100 mg/3 mL (52783-U)
condition: 1 mL methanol; 1 mL 10 mM potassium phosphate, pH 3
sample addition: 1 mL
washing: 1 mL 10 mM potassium phosphate, pH 3; 1 mL methanol
elution: 1 mL 5% ammonium hydroxide in methanol
eluate post-treatment: evaporate to dryness with nitrogen at room temperature, reconstitute in 1 mL mobile phase
column: Discovery HS F5, 15 cm x 4.6 mm I.D., 5 μm particles (567516-U)
mobile phase: [A] 10 mM potassium phosphate, pH 6: [B] acetonitrile (85:15, A:B)
flow rate: 2 mL/min
column temp.: 25 °C
detector: UV, 220 nm
injection: 10 μL
Compatibilità
application for HPLC
application for SPE
applicazioni
clinical
Applicazioni
Piroxicam and 2-aminopyridine (piroxicam′s polar metabolite) was spiked into human urine at the levels of 4 and 10 μg/mL, respectively. The urine sample was diluted 1:1 with 10 mM potassium phosphate, pH 3.
Note legali
Colonna analitica
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.