84-109
HPLC Analysis of Triazine Herbicides in Grass on SUPELCOSIL™ LC-8-DB after SPE using Supelclean™ LC-SCX
application for SPE, application for HPLC
About This Item
tecniche
HPLC: suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: fresh grass clippings spiked with herbicides at 2 ppm
SPE tube/cartridge: Supelclean LC-SCX, 500 mg/3 mL (57018)
condition: 1 mL methylene chloride
sample addition: 2 mL grass extract. Wash with 2 x 2 mL acetonitrile.
drying: Dry packing for 5 min. under nitrogen purge
washing: Dry packing for 5 min under nitrogen purge; wash with 2 x 2 mL DI water
elution: 1.5 mL methanol
eluate post-treatment: dilute to 2 mL with DI water
column: SUPELCOSIL LC-8-DB, 15 cm × 4.6 mm I.D., 5 μm (58347-U) with guard column (59553)
mobile phase: [A] acetonitrile; [B] water; (45:55, A:B)
flow rate: 1.5 mL/min
detector: UV, 254 nm
injection: 100 μL
Compatibilità
application for HPLC
application for SPE
applicazioni
environmental
Applicazioni
5 g fresh grass clippings spiked with 2 ppm each herbicide. Add 4 g anhydrous sodium sulfate and 20 mL methylene chloride:acetone (80:20). Shake 20 min and allow mixture to stand 1 min.
Note legali
Analita
Colonna analitica
Colonnina o piastra per SPE
Precolonna
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.