G006300
LC/MS Analysis of Aflatoxin M1 in Milk on Ascentis® Express HILIC after SPE using Supel™ Tox AflaZea, Quantifying Picogram Concentrations
application for SPE, application for HPLC
About This Item
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: cow milk
SPE tube/cartridge: Supel™ Tox AlflaZea, 6 mL (55314-U)
extraction process: centrifuge 45 g liquid milk for 60 min at 5000 rpm (RCF 4863 xg), cool at 0 °C for 30 minutes, remove fat layer, transfer middle layer to collection flask,
sample addition: 10 mL, 2 drop/second using vacuum
washing: 2 x 2 mL acetonitrile (collect with sample)
eluate post-treatment: evaporate to dryness at 70 °C using nitrogen
sample preparation: reconstitut in 500 μL 20:80 acetonitrile:water, filter into 750 μL poly vial using PVDF filter, 0.2 μm
column: Ascentis Express C18, 50 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
mobile phase: [A] 0.1% (v/v) formic acid; [B] acetonitrile
gradient: 90% A for 0.5 min; to 10% A in 0.5 min; held at 10% for 3 min; to 90% A in 0.10 min; held at 90% A for 3 min
flow rate: 0.4 mL/min
column temp.: ambient
detector: MS/MS (MRM transitions)
injection: 10 μL
suitability
application for HPLC
application for SPE
application(s)
food and beverages
Analysis Note
Legal Information
Analyte
SPE tube or plate
analytical column
related product
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service