材料

SPE管或板

产品编号
说明
价格

Discovery® DSC-Si SPE 小柱

bed wt. 1 g, volume 6 mL, pk of 30

分析柱

产品编号
说明
价格

Discovery® HS C18 高效液相色谱柱

5 μm particle size, L × I.D. 25 cm × 4.6 mm

標準

产品编号
说明
价格

甘油三油酸酯

≥99%

三亚油酸甘油酯

≥98% (TLC), liquid

橄榄油

analytical standard

CONDITIONS

SPE tube/cartridge

Discovery DSC-Si, 1 gm/6 mL (52656-U)

condition

6 mL hexane

sample addition

0.12 g oil in 0.5 mL hexane

elution

10 mL hexane-diethyl ether 87-13 (v/v)

eluate post-treatment

evaporation to dryness, reconstituted into 2 mL acetone

column

Discovery HS C18, 25 cm x 4.6 mm I.D., 5 μm particles (568523-U)

mobile phase

[A] acetone; [B] acetonitrile; (60:40, A:B)

flow rate

1.5 mL/min

pressure

914 psi (63 bar)

column temp.

30 °C

detector

ELSD

injection

5 μL

说明

分析報告

Olive oil was pre-cleaned using Silica SPE, analyzed as 5% sample in acetone (w/v).

法律資訊

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany