GC Analysis of Volatiles in Water on SPB®-1 after SPME using 100 μm PDMS Fiber

GC Analysis of Volatiles in Water on SPB®-1 after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

材料

固相微萃取纖維

产品编号
说明
价格

SPME 纤维组件聚二甲基硅氧烷 (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

CONDITIONS

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

5 min immersed in water with rapid stirring

column

SPB-1, 10 m × 0.20 mm I.D., 1.2 μm (24134-U)

oven

40 °C ( 0.75 min) to 160 °C at 20 °C/min

inj. temp.

230 °C, splitless (closed 3 min)

detector

FID, 260 °C

carrier gas

helium, 40 cm/sec

sample

50ppb each analyte in 1.8 mL saturated salt water