跳转至内容
Merck

GC Analysis of Cocaine in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

GC Analysis of Cocaine in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

材料

分析柱

产品编号
说明
价格

Equity-1 毛细管GC色谱柱

L × I.D. 30 m × 0.32 mm, df 0.25 μm

固相微萃取纖維

产品编号
说明
价格

SPME 纤维组件聚二甲基硅氧烷 (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

CONDITIONS

sample/matrix

0.5 mL urine (250 ng each analyte, 20 μL 2.5% NaF) in 1 mL vial

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

immersion, 30 min

desorption process

3 min, 240 °C

column

100% methyl polysiloxane, 30 m × 0.32 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28055-U)

oven

120 °C to 280 °C at 10 °C/min

inj. temp.

splitless (splitter opened after 1 min), 240 °C

detector

NPD, 280 °C

carrier gas

helium, 3 mL/min