G005827
LC/MS (TOF) Analysis of Warfarin Enantiomers in Plasma on Astec® CHIROBIOTIC® V after SPE using HybridSPE®-Phospholipid
application for HPLC
About This Item
Technique(s)
HPLC: suitable
Paramètres du test
sample/matrix: Rat plasma stabilized with K2EDTA was acquired from Lampire Biological Laboratories, (Pipersville PA). (Plasma was spiked directly from stock standard to a level of 100 ng/mL in 3:1 (plasma:1% formic acid acetonitrile).)
SPE tube/cartridge: Pk1 50 mg/96-Well HybridSPE-Phospholipid Plate (575656-U)
condition: Apply 100 μL of plasma to plate, followed by 300 μL of 1% formic acid acetonitrile. Agitate via vortex for 4 min.
elution: Place on vacuum manifold and apply 10” Hg vacuum for 4 minutes. Collect filtrate and analyze directly.
column: CHIROBIOTIC V,10 cm x 4.6 mm I.D., 5 μm particles (11022AST)
mobile phase: [A] 0.1% formic acid water; [B] acetonitrile, (75:25, A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: ESI+, 100-1000 m/z
injection: 2 μL
Adéquation
application for HPLC
Application(s)
clinical
pharmaceutical (small molecule)
Remarque sur l'analyse
Informations légales
Analyte
Colonne analytique
Tube ou plaque pour SPE
Choose from one of the most recent versions:
Certificats d'analyse (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique