G005599
LC/MS Analysis of Bile Acids and Their Conjugates on Ascentis® Express C18
application for HPLC
About This Item
Technique(s)
HPLC: suitable
Paramètres du test
sample/matrix: Plasma proteins were removed with addition of 900 μL of acetonitrile, containing deuterated internal standards, to 250 μL of human EDTA plasma. The mixture was vortexed, (centrifuged and the supermatant evaporated before being reconstituted in a 50:50 solution of methanol and water. 10 μL (corresponding to 8.57 μL of plasma) was injected into the HPLC.)
column: Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)
mobile phase: [A]: Water (5 mM ammonium acetate and 0.012% formic acid); [B]: Methanol (5 mM ammonium acetate and 0.012% formic acid)
gradient: 70 to 95% B in 10 minutes held 4 minutes
flow rate: 0.6 mL/min
column temp.: 40 °C
detector: ESI(-), MRM Mode
Adéquation
application for HPLC
Application(s)
clinical
Remarque sur l'analyse
Informations légales
Analyte
Colonne analytique
Composant de phase mobile
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique