Accéder au contenu
Merck
Toutes les photos(1)

Key Documents

KPHLS1THH1

Millipore

Durapore® 0.45 µm, Opticap® XLT Capsule

Opticap® XLT 10, cartridge nominal length 14.8 in. (37.6 cm), filter diam. 4.2 in. (10.7 cm)

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Code UNSPSC :
23151806

Matériaux

polyester support
polyethylene support
polypropylene filter
polypropylene housing
silicone seal

Niveau de qualité

Conformité réglementaire

FDA 21CFR177-182 (all component materials)

Stérilité

sterile; irradiated

Gamme de produits

Opticap® XLT 10

Caractéristiques

hydrophilic

Fabricant/nom de marque

Opticap®

Paramètres

≤15.0 mL/min air diffusion at 3.9 bar (56 psig) and 23 °C (in water)
1 bar max. inlet pressure (15 psi) at 80 °C
2.1 bar max. differential pressure (30 psid) at 4-40 °C (Reverse:)
2.8 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
4.1 bar max. differential pressure (60 psid) at 25 °C (Forward:)
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure

Technique(s)

bioburden reduction: suitable

Longueur nominale des cartouches

14.8 in. (37.6 cm)

device L

37.6 cm (14.8 in.)

device size

10 in.

Diam. du filtre

4.2 in. (10.7 cm)

filter filtration area

0.62 m2

Diam. du raccord d'entrée

5/8 in.

inlet to outlet width

37.6 cm (14.8 in.)

Diam. du raccord de sortie

5/8 in.

Impuretés

≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)

Matrice

Durapore®

Dimension de pores

0.45 μm pore size

Point de bulle

≥1930 mbar (28 psig), air with water at 23 °C

Raccords

1/4 in. fitting (vent/drain hose barb with double O-ring seal)
(16 mm (5?8 in.) Hose Barb Inlet and Outlet)

Vous recherchez des produits similaires ? Visite Guide de comparaison des produits

Description générale

Device Configuration: Capsule

Notes préparatoires

Sterilization Method:
Minimum sterilization up to 25 kGy. 3 autoclave cycles of 60min at 3°C; not in-line steam sterilizable

Remarque sur l'analyse

Gravimetric Extractables: ≤ 25 mg per capsule after 24 hours in water at controlled room temperature.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 1500 mL

Autres remarques

Directions for Use:

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer installation section of Opticap® XL Capsules and Opticap® XLT Capsules user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Informations légales

Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany

Vous ne trouvez pas le bon produit ?  

Essayez notre Outil de sélection de produits.


Certificats d'analyse (COA)

Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique