GC Analysis of Permanent Gases on Carboxen®-1010 PLOT

suitable for GC

Sign Into View Organizational & Contract Pricing


About This Item

technique(s)

gas chromatography (GC): suitable

test parameters

column: Carboxen-1010 PLOT, 30 m x 0.53 mm I.D. (25467)
oven: 35 °C (7.0 min), 24 °C/min to 225 °C
inj. temp.: 200 °C
detector: TCD, 230 °C
carrier gas: helium, 3.0 mL/min
injection: 10 μL, direct using a sample valve, 150 °C
sample: Oxygen, carbon monoxide, methane, and carbon dioxide, each at 1.0% in nitrogen