GC Analysis of Trichloroanisole (TCA) in Wine on SLB®-5ms after SPME using 100 μm PDMS Fiber
Materiales
Estándar
Referencia del producto
Descripción
Precios
Fibra para SPME
Referencia del producto
Descripción
Precios
Filtro SPME Polidimetilsiloxano (PDMS)
metal alloy fiber, df 100 μm(PDMS, needle size 23 ga, for use with autosamplercolumna analítica
Referencia del producto
Descripción
Precios
CONDITIONS
sample/matrix
red wine spiked with 5 ppt TCA
SPME fiber
100 μm PDMS metal fiber (57928-U)
extraction
headspace, room temp (30 min)
condition
3 min. at 250 °C
desorption process
250 °C for 3 min
column
SLB®-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
60 °C (2 min), 8 °C/min to 200 °C (10 min)
MSD interface
300 °C
scan range
SIM, m/z = 195, 197, 210, 212
carrier gas
helium, 1 mL/min. constant
liner
0.75 mm I.D. SPME
Descripción
Información legal
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany