LNEBIO302
Glucose, Creatinine, Glycerides, Cholesterol, Urea and Uric acid in human frozen serum
LNE Certified Reference Material
Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización
About This Item
Productos recomendados
formulario
liquid
caducidad
limited shelf life, expiry date on the label
composición
human serum, 100%
DIV
not for in vitro diagnostic use
temp. de almacenamiento
−70°C
¿Está buscando productos similares? Visita Guía de comparación de productos
Descripción general
LNE CRM Bio 302 is a higher-order certified reference material produced by the French metrological institute LNE in compliance with ISO 17034 requirements. This secondary standard is intended for use as quality control material to assess the measurement bias or measurement uncertainty of established or new measurement procedures for the determination of creatinine, glucose, total cholesterol, total glycerides, HDL-cholesterol and LDL-cholesterol in human serum.
The Certified Reference Material LNE CRM Bio 302 corresponds to a pool of Human Frozen Serum that was prepared according to the Clinical Laboratory Standards Institute (CLSI) C37-A guidelines describing the preparation and validation of commutable frozen human serum pools as Secondary Reference Materials for Cholesterol Measurement. Samples commutability was assessed through a commutability study involving several clinical laboratories. Commutability results are reported in the certificate of analysis.
Cert_ LNE CRM Bio 302
The Certified Reference Material LNE CRM Bio 302 corresponds to a pool of Human Frozen Serum that was prepared according to the Clinical Laboratory Standards Institute (CLSI) C37-A guidelines describing the preparation and validation of commutable frozen human serum pools as Secondary Reference Materials for Cholesterol Measurement. Samples commutability was assessed through a commutability study involving several clinical laboratories. Commutability results are reported in the certificate of analysis.
Cert_ LNE CRM Bio 302
Aplicación
LNE CRM Bio 302, a secondary standard, is intended for use as quality control material to assess the measurement bias or measurement uncertainty of established or new measurement procedures for the determination of creatinine, glucose, total cholesterol, total glycerides, HDL-cholesterol and LDL-cholesterol in human serum.
Características y beneficios
- A reliable certified reference materials (CRM) for clinically relevant biomarkers
- Concentrations of glucose, creatinine, total cholesterol, triglyceride (TG), high-density lipoprotein (HDL)-cholesterol, low-density lipoprotein (LDL)-cholesterol, urea, and uric acid, are certified with higher-order reference methods.
- Derived from real patient samples.
- This CRM is pooled, tested, and certified by the LNE, which includes extensive commutability studies.
- Commutability assessed for the main routine assays.
- The accuracy of the reference methods used to certify target values is verified using certified matrix reference materials.
- This CRM was used for national quality control organized by the ANSM in 2016 to assess the accuracy of some general biochemistry tests.
Otras notas
- Details on storage, safety, usage, and operating instructions are provided in the comprehensive Certificate of Analysis.
- Concentrations of glucose, total glycerides, and total cholesterol are certified by reference measurement procedures relying on gas chromatography isotope dilution mass spectrometry (GC-IDMS) listed in JCTLM database.
- Concentrations of HDLC and LDL-C, are certified at the CDC by beta quantification with cholesterol being determined by the Abell Kendall method.
- Concentrations of creatinine and uric acid are certified by reference measurement procedures relying on liquid chromatography isotope dilution mass spectrometry (LC-IDMS) listed in JCTLM database.
- This clinical matrix CRM is for research and clinical testing applications.
Código de clase de almacenamiento
10 - Combustible liquids
Clase de riesgo para el agua (WGK)
WGK 1
Punto de inflamabilidad (°F)
Not applicable
Punto de inflamabilidad (°C)
Not applicable
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
¿No ve la versión correcta?
Si necesita una versión concreta, puede buscar un certificado específico por el número de lote.
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico