G006163
HPLC Analysis of Exenatide Acetate on Proteomix® SCX-NP5
application for HPLC
About This Item
técnicas
HPLC: suitable
parámetros analíticos
column: Proteomix SCX NP5, 25 cm x 4.6mm I.D., 5 μm particles (Z777161)
mobile phase: (A) 10 mM monosodium phosphate:acetonitrile (56:44, v/v), adjusted to pH to 2.0 with phosphoric acid; ((B) 0.5 M NaCl/10 mM monosodium phosphate:acetonitrile (56:44, v/v), adjusted to pH to 2.0 with phosphoric acid)
gradient: 0 to 65% B in 35 min; held for 5 min; to 0% B in 10 min
flow rate: 1.0 mL/min
pressure: 1653 psi (114 bar)
column temp.: 40 °C
detector: UV, 235 nm
injection: 20 μL
sample: 0.325 mg/mL exenatide acetate in mobile phase A
idoneidad
application for HPLC
aplicaciones
life science and biopharma
pharmaceutical (small molecule)
Información legal
columna analítica
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico