G005921
GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis
suitable for GC, application for SPE
About This Item
Productos recomendados
técnicas
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
parámetros analíticos
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 10 g whole cow′s milk
extraction tube: Supel™ QuE Non-buffered tube (4 g magnesium sulfate, 1 g sodium chloride) (55294-U)
clean-up tube: Supel™ QuE Z-Sep 150 mg, magnesium sulfate (150 mg, part no. 63135) (55299-U)
column: SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 75 °C (1 min), 12 °C/min to 340 °C (10 min)
inj. temp.: 250 °C
detector: μ-ECD, 340 °C
carrier gas: hydrogen, 1.2 mL/min, constant
injection: 1 μL splitless (0.75 min)
liner: 4 mm I.D. FocusLiner™ with taper
idoneidad
application for SPE
aplicaciones
food and beverages
Descripción general
Nota de análisis
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL
Información legal
Estándar
columna analítica
tubo o placa de SPE
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico