Saltar al contenido
Merck
Todas las fotos(1)

Key Documents

G005621

GC Analysis of FAMEs in Cookies on SP®-2560 after SPE Fractionation using Discovery® Ag-Ion

suitable for GC, application for SPE

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

técnicas

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

parámetros analíticos

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 1 g of commercially purchased cookies was ground and subjected to acid digestion and alkali hydrolysis followed by methylation as described in AOCS Official Method Ce 1k-09
SPE tube/cartridge: Discovery Ag-Ion SPE tubes, 750 mg/6 mL (54225-U)
condition: 4 mL of acetone; allow solvent to gravity drip completely through tube; discard eluant; 4 mL of hexane; allow solvent to gravity drip completely through tube; discard eluant
sample addition: 1 mL of extract; discard any eluant that drips through tube
elution: (Fraction 1) 6 mL of hexane:acetone (96:4); collect eluant in a fresh container with slight vacuum; ((Fraction 2) 4 mL of hexane: acetone (90:10); collect eluant in a fresh container with slight vacuum;(Fraction 3) 4 mL of 100% acetone; collect eluant in a fresh container with slight vacuum)
eluate post-treatment: evaporate each fraction at room temperature using nitrogen; reconstitute each fraction to 1 mL of hexane
column: SP-2560, 100 m x 0.25 mm I.D., 0.20 μm (24056)
oven: 180 °C
inj. temp.: 250 °C
detector: FID, 250 °C
carrier gas: hydrogen, 1 mL/min
injection: 1 μL, 10:1 split
liner: 4 mm I.D., split type, single taper wool packed FocusLiner design

idoneidad

application for SPE

aplicaciones

food and beverages

Nota de análisis

A sample of commercially purchased cookies were prepared using an acid digestions /alkali hydrolysis followed by methylation as described in AOCS Official Method Ce 1k-09 (1). The resulting extracts were then fractionated using Ag-ion SPE and analyzed by GC/FID on the 100 m SP-2560 capillary column.

Información legal

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany
FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SP is a registered trademark of Sigma-Aldrich Co. LLC

Producto relacionado

Referencia del producto
Descripción
Precios

columna analítica

Referencia del producto
Descripción
Precios

tubo o placa de SPE

Referencia del producto
Descripción
Precios

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico