G003764
HPLC Analysis of Furosemide (Lasix) in Horse Serum on Discovery® C18 after SPE using Discovery® DSC-18
application for SPE, application for HPLC
About This Item
técnicas
HPLC: suitable
solid phase extraction (SPE): suitable
parámetros analíticos
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: acidified horse serum spiked with indapamide at 10 mg/mL
SPE tube/cartridge: Discovery DSC-18, 50 mg/1 mL (52601-U)
condition: 1 mL methanol; 1 mL 10 mM potassium phosphate, pH 3
sample addition: 1 mL
washing: 10 mM potassium phosphate, pH 3
elution: 1 mL 60% methanol in DI water.
column: Discovery C18, 15 cm x 4.6 mm I.D., 5 μm particles (504955)
mobile phase: [A] 10 mM KH2PO4, pH 3 (adjusted with H3PO4): [B] acetonitrile; (60:40, A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: UV, 234 nm
injection: 10 μL
idoneidad
application for HPLC
application for SPE
aplicaciones
forensics and toxicology
Aplicación
Neat horse serum was acidified with 10 mL 6 M HCl/mL serum. Internal standard (I.S.) blank was prepared by spiking acidified horse serum with indapamide at 10 mg/mL. Sample A = 10 mg/mL furosemide in I.S. blank. Sample B = 5 mg/mL furosemide in I.S. blank. Sample C = 0.5 mg/mL furosemide in I.S. blank. Sample D = 0.1 mg/mL furosemide in I.S. blank. Sample E = 0.05 mg/mL furosemide in I.S. blank.
Información legal
columna analítica
tubo o placa de SPE
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico