Multilayer Durapore® 0.45/0.22 µm, Opticap® XLT Capsule
Opticap® XLT 30, cartridge nominal length 30 in. (75 cm), filter diam. 4.2 in. (10.7 cm)
About This Item
Productos recomendados
Materiales
polyester support
polyethylene support
polypropylene filter
polypropylene housing
silicone seal
Nivel de calidad
cumplimiento norm.
FDA 21CFR177-182 (all component materials)
esterilidad
sterile; irradiated
Línea del producto
Opticap® XLT 30
Características
hydrophilic
fabricante / nombre comercial
Opticap®
Parámetros
≤45.0 mL/min air diffusion at 2.75 bar (40 psig) and 23 °C (in water)
1 bar max. inlet pressure (15 psi) at 80 °C
1.05 bar max. differential pressure (15 psid) at 80 °C (Forward:)
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward:)
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure
técnicas
sterile filtration: suitable
longitud nominal del cartucho
30 in. (75 cm)
device L
87.1 cm (34.3 in.)
device size
30 in.
diám. del filtro
4.2 in. (10.7 cm)
filter filtration area
0.1 m2
diám. conexión a entrada
5/8 in.
inlet to outlet width
19.8 cm (7.8 in.)
diám. conexión a salida
5/8 in.
impurezas
≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)
matriz
Multilayer Durapore®
pore size
0.45/0.22 μm pore size
punto de burbuja
≥3450 mbar (50 psig), air with water at 23 °C
Adaptador
1/4 in. fitting (vent/drain hose barb with double O-ring seal)
(16 mm (5?8 in.) Hose Barb Inlet and Outlet)
¿Está buscando productos similares? Visita Guía de comparación de productos
Descripción general
Envase
Nota de preparación
3 autoclave cycles of 60min at 126°C; not in-line steam sterilizable
This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).
Nota de análisis
Otras notas
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations
Información legal
¿No encuentra el producto adecuado?
Pruebe nuestro Herramienta de selección de productos.
Certificados de análisis (COA)
Busque Certificados de análisis (COA) introduciendo el número de lote del producto. Los números de lote se encuentran en la etiqueta del producto después de las palabras «Lot» o «Batch»
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico