US EPA Method 8260: GC Analysis of Volatiles on SPB®-624 after Purge & Trap using "K" Trap, Fast GC Analysis
Materialien
Analytische Säule
Probenvorb.
Purge/Trap K Vocarb® 3000
for use with Tekmar® Velocity XPT, LSC-1, LSC-2, LSC-2000, 4000 (Straight, 12 in./30.5 cm, Swagelok fitting), HT-3Purge/Trap K Vocarb® 3000
for use with Tekmar® 3000, 3100 (Straight, 12 in./30.5 cm, Valco fittingStandard
EPA 8260A Surrogat-Standardmischung
certified reference material, 2000 μg/mL each component in methanolEPA 8260 flüchtige Kalibriermischung
certified reference material, 2000 μg/mL each component in methanolTCL Volatiles Gemisch
certified reference material, 2000 μg/mL each component in methanol: water (9:1)Zubehör
Hamilton® Spritze, Serie 1000 GASTIGHT®, PTFE-Luer-Lock
1005TLL, PTFE Luer lock, volume 5 mL, needle size (not included)Ähnliches Produkt
EPA-Flaschen, weißer Schraubverschluss aus Polypropylen mit Mittelloch, vorgereinigt und vormontiert
volume 40 mL, amber glass vial, O.D. × H 29 mm × 82 mm, pkg of 72 eaCONDITIONS
sample/matrix
each analyte at 50 ppb in 5 mL water
purge trap
VOCARB 3000 K (24940-U)
purge
40 mL/min at 25 °C for 11 min
dry purge
2 min
desorption preheat
205 °C
desorption temp.
210 ºC for 1 min
desorption flow
150 mL/min
bake
260 °C for 10 min
transfer line/valve temp.
110 °C
column
SPB-624, 20 m x 0.18 mm I.D., 1.0 μm (28662-U)
oven
40 °C (1 min), 11 °C/min to 125 °C, 35 °C/min to 230 °C (2 min)
inj. temp.
150 °C
MSD interface
220 °C
scan range
m/z = 35-400
carrier gas
helium, 1.5 mL/min
injection
100:1 split
liner
0.75 mm I.D. SPME
Beschreibung
Hinweis zur Analyse
US EPA Method 8260 describes the analysis of volatile organic compounds in solid wastes and ground waters. This application illustrates the analysis of many compounds commonly analyzed by this method using purge and trap coupled to GC-MS. A short, narrow I.D. SPB-624 column was used to shorten the analysis time to under 11 minutes. A high desorb flow through the trap was necessary to prevent band broadening of the gaseous VOCs. This was accomplished by increasing the split ratio, thus increasing the total flow through the trap during desorb.
Rechtliche Hinweise