GC Analysis of Phthalate Esters in Corn Oil (Spiked) on SLB®-5ms after SPME with a 100 μm PDMS Fiber

Materialien
Analytische Säule
SPME-Faser
SPME-Faseranordnung Polydimethylsiloxan (PDMS)
df 7 μm(PDMS, needle size 24 ga, for use with manual holderSPME-Faseranordnung Polydimethylsiloxan (PDMS)
df 100 μm(PDMS, for use with autosampler, needle size 23 gaSPME-Faseranordnung Polydimethylsiloxan (PDMS)
df 100 μm(PDMS, needle size 24 ga, for use with manual holderStandard
Phthalsäure-bis-2-ethylhexylester
certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, SwitzerlandPhthalsäuredibutylester
certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, SwitzerlandDiisobutylphthalat
certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, SwitzerlandÄhnliches Produkt
Fläschchen, Schraubgewinde, Klarglas (nur Fläschchen)
volume 15 mL, clear glass vial, thread for 18-400, pkg of 100 eaScrew cap, polypropylene, with open center
black phenolic hole cap, for use with 15 mL vial with 18-400 thread, pkg of 100 eaSepta, tan PTFE/Silikon
white tan PTFE/silicone, diam. × thickness 16 mm × 0.090 in., for use with 15 mL vial, pkg of 100 eaCONDITIONS
sample/matrix
corn oil spiked with phthalate esters at 500 μg/kg
SPME fiber
100 μm PDMS (57300-U)
extraction
headspace, 95 °C (measured in headspace)
desorption process
260 °C for 4 min
column
F.S. CAP. SLB-5ms 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven
60 °C (1 min), 10 °C/min to 330 °C (10 min)
inj. temp.
260 °C
carrier gas
helium, 0.6 mL/min, constant
liner
0.75 mm I.D. SPME
sample
Corn oil
Beschreibung
Hinweis zur Analyse
In this application, SPME was used for the analysis of phthalate esters from oily samples, specifically the oil-based flavor packets included with ramen noodle kits. Spiked samples of corn oil and flavored oils included with ramen noodle kits were analyzed using high temperature SPME headspace extraction using a 100 μm PDMS fiber which was then directly desorbed into the GC/MS system. Calibration standards were prepared in corn oil, and spiked samples of chicken and beef oil-based flavors were analyzed. The SLB-5ms GC columns provide consistently low-bleed, inert, efficient, and durable separations.
Rechtliche Hinweise
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany