G005827
LC/MS (TOF) Analysis of Warfarin Enantiomers in Plasma on Astec® CHIROBIOTIC® V after SPE using HybridSPE®-Phospholipid
application for HPLC
About This Item
Methode(n)
HPLC: suitable
Testparameter
sample/matrix: Rat plasma stabilized with K2EDTA was acquired from Lampire Biological Laboratories, (Pipersville PA). (Plasma was spiked directly from stock standard to a level of 100 ng/mL in 3:1 (plasma:1% formic acid acetonitrile).)
SPE tube/cartridge: Pk1 50 mg/96-Well HybridSPE-Phospholipid Plate (575656-U)
condition: Apply 100 μL of plasma to plate, followed by 300 μL of 1% formic acid acetonitrile. Agitate via vortex for 4 min.
elution: Place on vacuum manifold and apply 10” Hg vacuum for 4 minutes. Collect filtrate and analyze directly.
column: CHIROBIOTIC V,10 cm x 4.6 mm I.D., 5 μm particles (11022AST)
mobile phase: [A] 0.1% formic acid water; [B] acetonitrile, (75:25, A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: ESI+, 100-1000 m/z
injection: 2 μL
Eignung
application for HPLC
Anwendung(en)
clinical
pharmaceutical (small molecule)
Hinweis zur Analyse
Rechtliche Hinweise
Analyt
Analytische Säule
SPE-Röhrchen oder -Platte
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
Leider sind derzeit keine COAs für dieses Produkt online verfügbar.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.