G005530
HPLC Analysis of Bisphenol A in Water on Ascentis® Express C18 after SPE using Supelclean™ ENVI-18
application for HPLC
About This Item
Methode(n)
HPLC: suitable
Testparameter
SPE tube/cartridge: Supelclean ENVI-18, 500 mg/6 mL glass SPE tube (54331-U)
condition: 1 mL 1% formic acid in acetonitrile, then 1 mL water
sample: 5 mL spiked water blank, concentration of 0.2 μg/mL
elution: 2 mL 1% formic acid in acetonitrile
eluate post-treatment: 1 mL aliquot of eluent, and evaporate/reconstitute to 0.5 mL in acetonitrile
column: Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase: [A] water; [B] acetonitrile; (60:40, A:B)
flow rate: 0.4 mL/min
pressure: 3270 psi (256 bar)
column temp.: 35 °C
detector: UV, 230 nm
injection: 1 μL
Eignung
application for HPLC
Anwendung(en)
environmental
Hinweis zur Analyse
Rechtliche Hinweise
Analyt
Analytische Säule
SPE-Röhrchen oder -Platte
Standard
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
Leider sind derzeit keine COAs für dieses Produkt online verfügbar.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.