Direkt zum Inhalt
Merck

1.09173

Millipore

MAS-100 Iso NT® Externer mikrobieller Luftkeimsammler

with USB, digital I/O and ProfiBus interfaces

Synonym(e):

MAS-100 Iso NT® Remote Microbial Air Sampler

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

UNSPSC-Code:
41103403
NACRES:
NA.74

Materialien

aluminum hardware

Agentur

according to ISO 14698

Leistungsmerkmale

installed

Verpackung

pkg of 1 kit

Hersteller/Markenname

MAS-100 Iso NT®

Anwendung(en)

air monitoring
pharmaceutical

Kompatibilität

for use with 90 mm Petri dishes

Lagertemp.

no temp limit

Suchen Sie nach ähnlichen Produkten? Aufrufen Leitfaden zum Produktvergleich

Allgemeine Beschreibung

The MAS-100 ISO NT® is designed to monitor the microbial contamination of the air in isolators. Like all other MAS-100 air samplers, it′s based on the Andersen impaction principle. The control unit with electronic and moving blower parts can be located outside the critical zone, while the single sampling head for standard 90 to 100mm Petri dishes collect samples in the containment zone. The tubing length can be up to 10 m when used with VHP decontamination.
The innovative double valve system enables the sampling heads to be integrated into the decontamination process with VHP. Each unit has both an air and a sterilization pump that operate autonomously. The instrument can be operated via an LCD screen.
The fully automatic calibration of the device guarantees maximum efficiency and accuracy. The system measures the air flow and regulates it to a constant value of 100 liters/min. These systems have the highest airflow accuracy of ±2.5%, compared to others that can be as high as ±10%. The instruments are designed and built according to GAMP 4 and are independently validated according to ISO 14698-1 with very good physical efficiency providing a D50 value below 1.1μm and a high biological efficiency result. The efficiencies and the D50 values are independent of the tube length and are comparable to the other members of the MAS-family.
The MAS-100 ISO NT®is equipped with one valve for one sampling head and RS232, USB and digital IN/OUT interfaces for communication. In case another communication interface is required MAS-100 ISO NT® Control unit with Ethernet with Ethernet interface is available. The type of sample base (standard or easy clean version) as well as the perforated lids can be selected separately according to the specific needs of the sample location and workflow.

Anwendung

MAS-100 ISO NT® Control unit with Profibus is suitable for microbial monitoring in isolators and cleanrooms. The microbial air sampling devices are extensively used in Pharmaceutical applications.

Komponenten

Control unit equipped with display and valve connection (including cable) for one sampling head, 2 Tri-Clamps, 2 Silicon gaskets, a range of phoenix connectors, Profibus module, USB cable, digital I/O, USB stick with all manuals and documentation.

Rechtliche Hinweise

MAS-100 ISO NT is a registered trademark of MBV AG

Zubehör

Produkt-Nr.
Beschreibung
Preisangaben

Ähnliches Produkt

Produkt-Nr.
Beschreibung
Preisangaben

Lagerklassenschlüssel

10-13 - German Storage Class 10 to 13


Analysenzertifikate (COA)

Suchen Sie nach Analysenzertifikate (COA), indem Sie die Lot-/Chargennummer des Produkts eingeben. Lot- und Chargennummern sind auf dem Produktetikett hinter den Wörtern ‘Lot’ oder ‘Batch’ (Lot oder Charge) zu finden.

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Verwandter Inhalt

Where production is subject to the most stringent requirements, our integrated microbial air monitoring systems can sample and monitor isolators and RABS for airborne organisms with the greatest accuracy.

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.