G006369
HPLC Analysis of PAHs in Canola Oil on SUPELCOSIL™ LC-PAH after SPE using Supelclean™ EZ-POP NP
application for SPE, application for HPLC
About This Item
Produits recommandés
Technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
Paramètres du test
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: Canola oil spiked with PAHs at 10 ng/g
SPE tube/cartridge: Supelclean™ EZ-POP NP, 12 mL, pk of 20 (54341-U)
condition: 10 mL acetone, dry under 10-15" mercury for 10 minutes
sample addition: 0.5 mL weighed accurately onto cartridge
elution: 15 mL acetonitrile
eluate post-treatment: concentrate to final volume of 0.5mL at 40 °C under nitrogen
column: SUPELCOSIL™ LC-PAH 25 cm x 4.6 mm I.D., 5 μm particles (58229)
mobile phase: [A] water [B] acetonitrile
gradient: 40% B held for 5 min; to 100% B in 15 min; held at 100% B for 12 min
flow rate: 2.4 mL/min
pressure: 2790 psi at start (192 bar)
column temp.: 25 °C
detector: Fluorescence, programmed.
Adéquation
application for HPLC
application for SPE
Application(s)
food and beverages
Remarque sur l'analyse
Autres remarques
Informations légales
Analyte
Colonne analytique
Tube ou plaque pour SPE
Étalon
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique