GC Analysis of Pesticides in Beef Kidney on SLB®-5ms (20 m x 0.18 mm I.D., 0.36 μm) after QuEChERS Cleanup using Supel Que Z-Sep+, MS in SIM Mode, Fast GC Analysis

suitable for GC, application for SPE

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

Paramètres du test

sample preparation: Dispersive (QuEChERS)
sample/matrix: 10 gm homogenized beef kidney spiked at 50 ng/g with pesticides
extraction tube: Supel Que Acetate (AC) Tube (55234-U)
extraction process: add 10 mL of acetonitrile; shake for 1 minute; add contents of Supel Que acetate extraction tube (55234-U); shake immediately for 1 minute; centrifuge at 3200 rpm for 5 minutes; (Sample Pretreatment: Homogenize 10 gm beef kidney and spike with pesticides at 50 ng/g.)
clean-up tube: Supel QuE Z-Sep+ tube (55296-U)
clean-up process: transfer 3 mL of the acetonitrile layer into a Supel QuE Z-Sep+ cleanup tube (55296-U); shake for 1 minute; centrifuge at 3400 rpm for 3 minutes; draw off supernatant for LVI-GCMS analysis
column: F.S. CAP. SLB-5MS 20 m x 0.18 mm I.D., 0.36 μm (28576-U)
oven: 70 °C ( 0.5 min), 25 °C/min to 125 °C, 10 °C/min to 200 °C, 5 °C/min to 300 °C/min (1 min)
inj. temp.: programmed, 60 °C (0.28 min), 600 °C/min to 325 °C (5 min)
detector: MSD
MSD interface: 325 °C
carrier gas: helium, 1 mL/min constant flow
injection: 10 μL LVI, PTV solvent vent, rapid injection speed
liner: 4 mm ID FocusLiner with taper
sample: Beef Kidney extract

Adéquation

application for SPE

Application(s)

food and beverages

Remarque sur l'analyse

Beef kidney was analyzed for pesticides by GC/MS-SIM after extraction and cleanup. Dispersive SPE was used for both extraction and cleanup. Since beef kideny contains approximately 5% fat, Z-Sep + was used for cleanup.

Informations légales

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC

Analyte

Description

    Diphenylamine
    Hexachlorobenzene
    Lindane
    Endosulfan I
    4,4′-DDE
    1,1-Dichloro-2,2-bis(4-chlorophenyl)ethane
    Endosulfan sulfate
    4,4′-DDT
    Piperonylbutoxide
    Bifenthrin
    Permethrin (isomers) solution
    Afficher tout (11)

Tube ou plaque pour SPE


Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique