G005535
GC Analysis of Volatiles in Green Arabica Coffee on SUPELCOWAX 10 after SPME using 50/30 μm DVB/Carboxen/PDMS Fiber
suitable for GC, application for SPME
About This Item
Technique(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
Paramètres du test
sample/matrix: Arabica coffee beans, green, 2 grams with pre-equilibration of 10 min at 60 °C, the entire process under agitation at 500 rpm
SPME fiber: divinylbenzene/carboxen on polydimethylsiloxane, 50/30 μm (57328-U)
extraction: headspace, 40 min at 60 °C
desorption process: 5 min
column: SUPELCOWAX 10, 30 m × 0.25 mm I.D., 0.25 μm (24079)
oven: 40 °C (5 min), 4 °C/min to 230 °C, 50 °C/min to 280°C (2 min)
inj. temp.: 260 °C
detector: MS (Shimadzu GCMS-QP2010) ionization energy 1.5 kV; mass range: 40-400 amu; scan mode, scan interval : 0.5 sec
MSD interface: 230 °C, source temp. 200 °C
carrier gas: helium, 34.0 cm/s and 40.3 KPa
injection: splitless 5 min, then split ratio 1:20
liner: 0.75 mm i.d.
Adéquation
application for SPME
Application(s)
food and beverages
Informations légales
Analyte
Colonne analytique
Fibre SPME
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique