G003893
GC Analysis of Closely Eluting Analytes from US EPA Method 8270 on SLB®-5ms (15 m x 0.10 mm I.D., 0.10 μm)
suitable for GC
About This Item
Methode(n)
gas chromatography (GC): suitable
Testparameter
column: SLB-5ms, 15 m x 0.10 mm I.D., 0.10 μm (28466-U)
oven: 40 °C (2 min), 22 °C/min to 240 °C, 10 °C/min to 330 °C (1 min)
inj. temp.: 250 °C
detector: FID, 330 °C
carrier gas: helium, 30 cm/sec @ 200 °C, set using methane
injection: 0.1 μL splitless (0.75 min)
liner: 2 mm I.D., straight
sample: 50 ng on-column of a 72 component semivolatile standard and 8 surrogate compounds, plus 6 internal standards (at 40 ng on-column), in methylene chloride