Millipore

Polysep II 2.0/1.2 µm nominal, Cartridge

cartridge nominal length 40 in. (100 cm), filter diam. 2.8 in.

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

UNSPSC-Code:
23151806

Materialien

borosilicate glass fiber (BGF) membrane
mixed cellulose esters (MCE) membrane
polypropylene device
polypropylene support (spun bonded)
silicone seal

Qualitätsniveau

Einhaltung gesetzlicher Vorschriften

meets FDA 21 CFR 177–182 (all component materials)

Sterilität

non-sterile

Kompatibilität mit Sterilisationsverfahren

autoclavable compatible
steam-in-place compatible

Produktlinie

Polysep

Leistungsmerkmale

hydrophilic

Hersteller/Markenname

Polysep II

Parameter

3.5 bar max. differential pressure (50 psid) at 25 °C

Methode(n)

prefiltration: suitable

Länge

40 in.

Breite

2.7 in.

Nennlänge Kartusche

40 in. (100 cm)

device size

40 in.

Filterdurchm.

2.8 in.

filter filtration area

1.84 m2

Verunreinigungen

≤0.5 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

Matrix

Polysep II

pore size

2.0/1.2 μm nominal pore size
2.0/1.2 μm pore size

Kartuschencode

Code 0

Suchen Sie nach ähnlichen Produkten? Aufrufen Leitfaden zum Produktvergleich

Allgemeine Beschreibung

Device Configuration: Cartridge
Cartridge Type: Liquid
Filter Grade: Prefilter

Anwendung

mAb & r-protein intermediates; Plasma; Vaccines & viral vectors; SVPs LVPs; Cell Culture; Media Filtration; Colloid Removal; Lipid Removal

Leistungsmerkmale und Vorteile

Format: Double Layer

Verpackung

Double Easy-Open bag

Angaben zur Herstellung

Sanitization: 30 min with hot water at 80 °C
Sterilization Method: 10 autoclave cycles or SIP at 121 °C for 30 min
This product was manufactured with a Millipore® media combination which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Hinweis zur Analyse

After autoclaving and a 24 hour soak in ASTM® Type 1 reagent grade water at controlled room temperature, extractables level was ≤140 mg per 10-inch element after a 5000 mL flush.
Cartridges meet the requirements of the USP Oxidizable Substances Test after a water flush of 5000 mL per 10 inch Element

Sonstige Hinweise

Directions for Use:
  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer first page of Visual Inspection Guide and introduction section of Mixed cellulose esters Tech Brief
  • Storage Statement: Please refer Mixed cellulose esters Tech Brief
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

The product is manufuactured in a facility with a Quality Management System approved by an accredited registering body to an ISO 9001 Quality Systems Standard.

Rechtliche Hinweise

ASTM is a registered trademark of American Society for Testing and Materials
Millipore is a registered trademark of Merck KGaA, Darmstadt, Germany
Polysep is a trademark of Sigma-Aldrich Co. LLC

Sie haben nicht das passende Produkt gefunden?  

Probieren Sie unser Produkt-Auswahlhilfe. aus.


Analysenzertifikate (COA)

Suchen Sie nach Analysenzertifikate (COA), indem Sie die Lot-/Chargennummer des Produkts eingeben. Lot- und Chargennummern sind auf dem Produktetikett hinter den Wörtern ‘Lot’ oder ‘Batch’ (Lot oder Charge) zu finden.

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.