HPLC Analysis of Paclitaxel and Related Compounds on Ascentis® Express F5
application for HPLC
About This Item
Technique(s)
HPLC: suitable
Paramètres du test
column: Ascentis Express F5, 15 cm x 4.6 mm I.D., 5 μm particles (50631-U)
mobile phase: [A] water; [B] acetonitrile; (60:40, A:B)
flow rate: 1.5 mL/min
pressure: 2727 psi (188 bar)
column temp.: 30 °C
detector: UV, 227 nm
injection: 5 μL
sample: taxols 25 μg/mL in 70:15:15, water:methanol:acetonitrile
Adéquation
application for HPLC
Application(s)
pharmaceutical (small molecule)
Remarque sur l'analyse
The developed method can be easily used for the routine quality control of bulk and Parenterals dosage form of Paclitaxel within a short analysis time. This method offers the advantages of higher efficies, reduced backpressure and shorter run times over a traditionally run methods over 5 μm totally porous particle columns.