84-109
HPLC Analysis of Triazine Herbicides in Grass on SUPELCOSIL™ LC-8-DB after SPE using Supelclean™ LC-SCX
application for SPE, application for HPLC
About This Item
Technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
Paramètres du test
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: fresh grass clippings spiked with herbicides at 2 ppm
SPE tube/cartridge: Supelclean LC-SCX, 500 mg/3 mL (57018)
condition: 1 mL methylene chloride
sample addition: 2 mL grass extract. Wash with 2 x 2 mL acetonitrile.
drying: Dry packing for 5 min. under nitrogen purge
washing: Dry packing for 5 min under nitrogen purge; wash with 2 x 2 mL DI water
elution: 1.5 mL methanol
eluate post-treatment: dilute to 2 mL with DI water
column: SUPELCOSIL LC-8-DB, 15 cm × 4.6 mm I.D., 5 μm (58347-U) with guard column (59553)
mobile phase: [A] acetonitrile; [B] water; (45:55, A:B)
flow rate: 1.5 mL/min
detector: UV, 254 nm
injection: 100 μL
Adéquation
application for HPLC
application for SPE
Application(s)
environmental
Application
5 g fresh grass clippings spiked with 2 ppm each herbicide. Add 4 g anhydrous sodium sulfate and 20 mL methylene chloride:acetone (80:20). Shake 20 min and allow mixture to stand 1 min.
Informations légales
Analyte
Colonne analytique
Colonne de garde
Tube ou plaque pour SPE
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique